Set controls as follows:
Main power switch: OFF
Hi voltage switch: OFF
Voltage control: 0
Process Control: auto or manual
Vent valve: closed
Pressure control: decrease
Mode selector: plate/DC

  • Open chamber by pushing slightly backward on glass cylinder to break suction at bottom and lifting glass cylinder.
  • Place cylinder upside down while holding top.
  • Place specimens on pedestal.
  • Replace glass cylinder and top plate.
  • Turn on main power switch.
  • Allow to pump down until vacuum reaches 20 millitorr or less (5-20 min when first turned on; 5-10 min thereafter).
  • Increase pressure control until vacuum reads 80-100 millitorr.
  • Maintain vacuum at 80-100 millitorr throughout coating process by adjusting pressure control.
  • If in auto, set timer to desired number of minutes by turning past 3 and then to desired position.
  • If in manual, set timer for desired time. Turn on hi voltage switch.
  • Turn volts control to obtain reading of 15 milliamperes.
  • When time has elapsed, turn volts knob to 0.
  • Turn hi voltage switch to off.
  • Turn pressure control to decrease.
  • Turn main power switch to off.
  • Turn vent valve to atm. Lift glass cylinder and top plate as before and remove specimen.
  • Replace glass cylinder and top plate.